Long-term Hair Removal Using a Pulsed, High-Power Diode Laser
Christine C. Dierickx, MD(l); R. Rox Anderson, MD (l);
Valeria B. Campos, MD(l);
Melanie C. Grossman, MD(2)
Wellman Labs of Photomedicine. Harvard Medical School
Laser and Skin Surgery Center of New York
Summary
provided by Lumenis,
Ltd.
The study summarized below was presented in progress to FDA. which cleared
the LightSheerTM Diode Laser System for hair removal in December 1997. This
summary is intended to allow users of this system to better inform their patients.
This is not a peer-reviewed medical publication, and is provided by Coherent
Medical solely for informational purposes regarding its products. It is not
a substitute for clinical observation of laser-tissue interaction and clinical
experience. Training is recommended prior to using the LightSheer Diode Laser
System.
INTRODUCTION
A solid-state, 800 nm pulsed near-infrared diode laser
was studied for long-term hair removal. The effect of laser fluence (energy
per unit area), single vs. multiple treatments, and single vs. multiple pulses
were determined in different skin types (Fitzpatrick's type I through VI).
Semiconductor diode lasers are considered the most efficient light sources
available and are particularly well suited for clinical applications. The
pulsed diode laser used in the study delivers high-energy laser pulses, in
combination with a proprietary skin cooling system, to target pigmented hair
follicles deep within the dermis. Treatment operates on the principle of selective
photothermolysis, which combines selective absorption of light energy by the
melanin in hair follicles with suitable pulse energies and pulse widths (pulse
duration) that are equal to or less than the thermal relaxation time (TRT)
of targeted follicles in human skin. There are two important anatomical targets
for inactivation of hair follicles: 1) stem cells in a "bulge" of the outer
root sheath about 1 mm below the skin surface; and 2) the dermal papilla located
at the deepest part of the follicle, which varies with hair growth cycle.
Research and extensive clinical use of lasers for hair removal have identified
important parameters to optimize the efficacy and safety of laser treatment:
Wavelength:
Most laser hair removal systems are designed to remove unwanted hair through
selective photothermolysis. This process involves local selective absorption
of an intense light pulse at wavelengths that: 1) are preferentially absorbed
by the desired hair follicles but not by the surrounding tissue; and 2) penetrate
deeply into the skin to reach the important targets for inactivation of hair
follicles. In laser hair removal, the most important and dominant absorber
is melanin. In all colors but white hair, there is sufficient melanin in the
follicular epithelium and matrix to act as a chromophore for light absorption
in the follicle. Laser energy is selectively absorbed by the melanin and causes
thermal damage to the hair shaft and follicle. Hair growth is impeded or eliminated
with sufficient fluence of the appropriate wavelength, due to selective thermal
damage of the hair follicle. The ideal laser wavelength for hair removal is
strongly absorbed by melanin but not by surrounding tissue and reaches deeply
into the dermis. Wavelengths between about 700 and 1000 nm fit these criteria.
. Pulse Width: Pulse width is a very important parameter for effective laser
hair removal without epidermal injury. For hair removal, the optimum pulse
duration is approximately equal to the thermal relaxation time (TRT) of the
hair follicle. The TRT is defined as the time required for an object to cool
to half the temperature achieved immediately following laser exposure. For
human terminal hair, TRT varies from about 10 to 100 milliseconds. Laser pulses
much shorter than the TRT cause insufficient heating of the target structures
(bulb and papilla) surrounding the hair shaft. Pulse widths much longer than
the TRT may cause non-selective damage to the surrounding dermis. The first
laser hair removal treatment to be cleared by FDA used Nd:YAG laser pulses
about a million times shorter than TRT for hair follicles, and failed to produce
long-term hair removal. The pulsed diode laser used in this study was specifically
designed to produce pulse widths matching the TRT of terminal hair follicles.
Fluence:
Previous studies have shown that stronger laser treatments,
or treatments using the highest tolerable fluence, produce better hair reduction
results. The risk of side effects also increases with fluence. In the study
summarized here, a range of fluences was given without regard to the skin
type of the patient.
Cooling:
Even laser light with perfect specificity for melanin
can cause damage to the skin surrounding the hair follicles because the epidermis
also contains melanin. Therefore, it is imperative to use an epidermal cooling
strategy to cool the epidermis while sufficient laser energy is delivered
to damage hair follicles. The most effective cooling method available is active
cooling. When in contact with a cold object, heat flows from the epidermis.
The important targets for hair removal lie at least 1 mm below the skin surface.
When the skin is actively cooled for 0.2-1 seconds, these targets remain warm,
while the epidermal temperature plummets. Epidermal temperatures less than
about -10 C cause tissue injury from freezing. Clinically, it is valuable
to cool the skin before, during and after the laser pulse for maximum epidermal
protection and patient comfort. Fast cooling requires good contact with a
cold, thermally conductive substance. Sapphire is ideal, as it has excellent
thermal characteristics and operates as a heat sink removing heat from the
epidermis. In addition, the diode handpick utilized in the study allowed compression
of the area being treated. Compression forms excellent thermal contact, collapses
blood vessels (a competing target), and forces hair to lie down, bringing
hair follicle roots closer to the surface. Consequently, the laser energy
is more effectively targeted to the intended site of action.
Number of Treatments:
Temporary hair removal is easily achieved in a single
treatment. The amount of long-term hair removal per treatment varies between
patients, increasing with the fluence used for treatment. Most patients require
more than one treatment, typically 2-5, to achieve nearly complete, long-term
hair Joss. In this study, 89% of the patients achieved significant long-term
hair reduction (defined as greater than 85% hair reduction) after one or two
treatments with the pulsed diode laser. 2 .
Number of Pulses (Single versus Multiple Pulsing):
Multiple pulses given to a site do not have significantly
greater effectiveness than a single pulse. However, the risk of pigmentary
side effects is somewhat increased. intentional multiple pulsing to a single
site should be avoided.
STUDY DESIGN
The primary objective was to investigate effectiveness
and safety of a pulsed diode laser in long-term removal of pigmented hair.
This large, long-term, prospective, blinded, controlled and quantitative study
was designed to study fluence-response relationship, one versus two treatments,
and single versus multiple pulses. Ninety-two (92) patients were treated at
two facilities: 46 patients at the Massachusetts General Hospital in Boston
and 46 at the Laser and Skin Surgery Center of New York, in New York City.
There were 45 males and 47 females with varying hair colors and skin types
(Fitzpatrick's skin type I to VI; predominately II to III). All patients were
treated and examined at 0, 1,3,6 and 9 months, and thirty-five patients were
also followed up at 12 months.
TABLE 1. FITZPATRICK CLASSIFICATION OF SKIN TYPES
Skin Type Characteristic:
I Always burns, never tans
II Always burns, sometimes tans
III Sometimes burns, always tans
IV Rarely burns, always tans
V Moderately pigmented
VI Black skin
The device used was a semiconductor diode laser system
that delivers pulsed, infrared light at a wavelength of 800 nm, pulse duration
from 5-20 ms and fluences from 15-40 J/cm2. Testing with diode lasers has
shown that at 800 NM, the laser light effectively penetrates the dermis, where
follicular melanin is the dominant chromophore. Given that the thermal relaxation
time for hair follicles ranges from J 0-100 ms, the pulse duration of 5-20
ms produced by this device is long enough to allow heat conduction from the
pigmented hair shaft during each pulse.
The laser hand piece contains high-power diode arrays, eliminating the need
for an articulated arm or fiber-optic beam delivery system. The hand piece
integrates a condenser that mixes light to produce a fluence of 15-40 J/cm2
over a uniform 9x9 mm area. The handpiece contains an actively cooled convex
sapphire lens that, when pressed against the patient's skin slightly before
and during each laser pulse, provides thermal protection for the epidermis.
The cooling lens not only allows higher doses of laser energy to safely and
effectively target hair follicles, but also allows compression of the target
area placing hair roots closer to the laser energy.
Before each treatment, eight test sites were positioned
on a patients.
An independent statistician performed data analysis.
Hair reduction was defined as the percentage of terminal hairs absent after
treatment, compared with the number before treatment. Hair reduction was quantified
at each follow-up visit tar each site, and the mean hair loss and standard
error were calculated.
RESULTS
Treatment demonstrated two different effects on
hair growth: hair growth delay and long- term hair reduction. A measurable
growth delay was seen in all patients ( 100%) at all fluence/pulse width configurations
tested; this growth delay was sustained for 1-3 months.
Table 4 shows percentage of hair reduction for all sites for all laser configurations.
After two treatments at 40 J/cm2 (20 ms pulse duration)3, the average long-term
hair reduction was 46%. Two treatments significantly increased hair reduction
as compared to one treatment, with an apparently additive effect. At a fluence
of 40 J/cm2, the initial treatment removed approximately 30% of terminal hairs,
and the second treatment given one month later removed an additional 25%.
Triple-pulsing (3x) did not significantly increase hair reduction over single
pulsing, after one or two treatments. However, the incidence of side effects
was higher for triple pulsing.
TABLE 4. HAIR REDUCTION RESULTS
Fluence Number of Treatments Percentage of Hair Reduction
1 mo 3 mo 6 mo. 9 mo. 12 mo 5 ms, 15 J/cm2 1 65.4 21.5 17.9* 15.5* 26.6 10
ms, 20 J/cm2 1 66.7 21.0 22.2 20.7 25.9 15 ms, 30 J/cm2 1 70.8 30.2 28.7 30.6
29.4 20 ms, 40 J/cm2 1 70.2 26.8 29.8 32.5 32.5 20 ms, 40 J/cm2 2 69.3 51.5
37.1 42.3 46.6 20 ms, 40 J/cm2 3x 2 71.1 51.9 36.8 41.4 46.2 20 ms, 40 J/cm2
3x 1 68.9 30.8 32.3 32.4 38.5 Control 0 17.3 10.5 10.8 6.3 5.5
Hair regrowth stabilized at 6 months at all fluences; there was no further
hair regrowth between 6, 9 and 12 months. This stabilizing of hair regrowth
or hair count is consistent with the clinically accepted growth cycle of follicles
(Table 5) and the definition of long-term term hair reduction, being a significant
reduction in the number of terminal hairs after treatment, which is stable
for a longer period than the complete growth cycle of follicles at the body
site tested.
Statistically significant reduction in average hair regrowth (p<0.01) continued
at 3,6,9, and 12 months for all sites, at all fluence-pulsewidth configurations,
after both one and two treatments. Eighty-nine percent of patients exhibited
significant long-term hair reduction at all configurations.
TABLE
5. DURATION OF GROWTH CYCLES
Location Telogen (months) Anagen (months) Total (months)
Back 3-6 3-6 6-12 Thigh 3-6 3-6 6-12 Arm 3-5 1-2 4-7 Calf 3-4 4-5 7-9
In addition to statistically significant hair reduction, treatment with the
laser also showed reduction in hair diameter and reduction in color of regrowing
hairs. Regrowing mean hair diameter decreased by 19.9%, and optical transmission
at 700 nm of hair shafts regrown post-treatment was 1.4 times greater than
transmission pretreatment (p<0.05). These added benefits of the treatment
are cosmetically desirable, since thinner, lighter hairs add to the appearance
of hair reduction.
Histological analysis suggested two mechanisms for effective, long-term removal
of terminal hair: miniaturization of coarse hair follicles to vellus-like
hair follicles, and destruction of the follicle with granulomatous degeneration
with a fibrotic remnant. The histological examination in this study showed
that treatments with the pulsed diode laser caused immediate thermal damage
in follicles with large, pigmented shafts, while follicles with small vellus
shafts showed no effect. Both pigmented and non-pigmented areas of terminal
hair follicle epithelium showed thermal coagulation necrosis, with minimal
or no damage to the adjacent dermis. Histological analysis also demonstrated
that triple pulsing did not produce more follicular damage than single pulsing,
although the dermis between closely spaced follicles was occasionally injured
by triple-pulsing. Sebaceous glands near the treated follicles showed no or
minimal thermal damage, rind sweat glands and dermal capillaries appeared
normal. This study was intended to elicit side effects, by covering a wide
range of fluences, regardless of skin type. Side effects with pulsed diode
laser treatment were fluence and skin type dependent. Hyper- or hypopigmentation
was minimal in fair skin, and increased " with fluence and with darker skin
type. At the highest fluence given of 40 J/cm2, the incidence of hyper- or
hypopigmentation was greater for patients with skin types III through VI.
In addition, clinical experience has shown that these high fluences may elicit
somewhat greater side effects in treatments of large areas. Immediately after
treatment, the typical response is perifollicular erythema and edema, which
subsides within a few hours. In this dose response study, all fluences were
given to most patients, regardless of skin type. (At the New York site, fluences
at or above those that showed evidence of epidermal injury were not delivered.
This resulted in several patients who did not receive the highest fluences.)
Approximately 20% of patients exhibited pigment changes which resolved in
1-3 months. The vast majority of pigment changes were transient, but with
darker skin types and higher fluences, some persistent pigment changes were
noted. Triple pulsing increased the incidence of hyper- or hypopigmentation
as compared to single pulsing, but did not significantly increase hair reduction.
CONCLUSIONS
The pulsed diode laser utilized in this study provides
a safe and effective treatment that achieves both temporary and long-term
reduction of unwanted, pigmented hair. Long-term hair reduction occurred in
89% of the patients in this large long-term, prospective, blinded, controlled
and quantitative study. On average, about half of the hair had long-term removal
after two treatments at a fluence of 40 J/cm2. Many patients had nearly complete,
long-term hair reduction after two treatments, while a few had little or no
long-term hair reduction. Regrowing hair is typically thinner and lighter
in color, adding to the cosmetic benefit. Both the efficacy for hair removal
and the risk of side effects increase with increasing treatment fluence. There
was an apparent threshold fluence for inducing side effects in each skin type.
The mechanisms for long-term hair reduction include miniaturization of terminal
hairs, and degeneration of follicles damaged by selective photothermolysis.
These study results support the clinical utility of the high-power, pulsed
diode laser as a safe and effective device for long-term removal of pigmented,
terminal hair. In clinical practice, fluence and pulse width should be adjusted
for skin type. At one clinical location over 1,000 clinical treatments were
performed with this de'v'ice, in which fluence and skin type were matched
to optimize the efficacy and safety of treatment. When this was done, the
incidence of side effects was less than 1%, and was limited to transient changes
in skin pigmentation.
Footnotes:
1 Now commercially available as the LightSheer Diode Laser System.
2 Pulse width setting for the system tested. Current system has a pulse width
setting of 7.5 ms for 15 J/cm2.
3 A lower fluence is recommended until significant experience is obtained
with the LightSheer. Please contact your Coherent sales representative
for a copy of Recommended Guidelines for Treating Patients for
more information.
RECOMMENDED REFERENCES
Absten GT, Joffe SN. Lasers in Medicine. An Introductory
Guide. Second Edition, Chapman and Halll:London, 1988.
Anderson RR "Laser Tissue Interactions", chapter 1 in: Cutaneous laser surgery:
the art and science of selective photothermolysis (Goldman MP, Fitzpatrick
RE, eds). St Louis: Mosby-Year Book, 2nd Ed., 1998: 1-18.